| Model Number |
391018 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The orsiro drug-eluting stent system was chosen for the treatment of a calcified lesion in the rca.After pre-dilatation the orsiro was delivered into the target lesion but resistance was felt.Upon removal a guideplus was inserted and the orsiro was re-inserted but there was still some resistance.Thus the orsiro was removed from the patients body and a stent deformation was detected.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.It should be noted that the ifu further the user warns against re-insertion if the stent system was removed prior to expansion.
|
| |
|