• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 3/80/130; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 3/80/130; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 357503
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/02/2020
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-18 balloon catheter was chosen for treatment.During inflation the balloon ruptured at 13 atm.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 19 mm.Microscopic inspection of the balloon surface showed deep scratches in close vicinity of the tear.It therefore seems likely that the tear in the balloon was caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations no manufacturing related root cause could be determined.The root cause is most likely related to the patients anatomy.It should be noted that, according to the ifu, the passeo-18 peripheral dilatation catheter is indicated to dilate stenosis in the femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 3/80/130
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10051163
Report Number1028232-2020-02009
Device Sequence Number835773
Product Code LIT
UDI-Device Identifier07640119559148
UDI-Public07640119559148
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number357503
Device Catalogue NumberSEE MODEL NO.
Device Lot Number02192042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Initial Date Received by Manufacturer 05/08/2020
Supplement Date Received by Manufacturer07/27/2020
Initial Report FDA Received Date05/13/2020
Supplement Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-