• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER; SURESTEP TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER; SURESTEP TRAY Back to Search Results
Model Number A119214M
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the cord on the temperature sensor had eroded through the shaft of the foley, near the 3-way port.It was noticed when the nurse was assessing the patient.It was immediately removed and there was no harm was done to the patient.No medical intervention was reported.
 
Event or Problem Description
It was reported that the cord on the temperature sensor had eroded through the shaft of the foley, near the 3-way port.It was noticed when the nurse was assessing the patient.It was immediately removed and there was no harm was done to the patient.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was confirmed.Visual evaluation of the returned sample noted one opened (no original packaging), used silicone foley.It was noted that at some points on the shaft near the trifurcation, the wire was irregularly coiled and penetrated and protruded out from the shaft.This did not meet the specification "the wire should not be kinked, knotted or broken once is inserted in the lumen".Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be short shaft/long thermistor.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER
Common Device Name
SURESTEP TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10051316
Report Number1018233-2020-03167
Device Sequence Number10956733
Product Code EZL
UDI-Device Identifier00801741073670
UDI-Public(01)00801741073670
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model NumberA119214M
Device Catalogue Number119214M
Device Lot NumberNGDW2512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2020
Initial Date Received by Manufacturer 04/22/2020
Supplement Date Received by Manufacturer06/16/2020
Initial Report FDA Received Date05/13/2020
Supplement Report FDA Received Date07/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-