| Model Number |
F2911-05550 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
04/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an everest fenestrated polyaxial screw fractured post-operatively.The patient did not fuse and revision surgery was performed to remove the fractured screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Functional, dimensional, and material analysis could not be performed as the device was not returned.However, x-rays were provided by the rep.Upon review, it was observed that a screw had fractured at the most caudal level of a construct.The fracture occurs approximately one centimeter below the proximal screw housing.A review of the device and complaint history records could not be performed as a valid lot code was not provided and could not be obtained.It is possible non-union contributed to the failure, as the patient did not fuse.However, as the device was not available for evaluation, the root cause could not be determined conclusively.H3 other text : status and location of the device is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that an everest fenestrated polyaxial screw fractured post-operatively.The patient did not fuse and revision surgery was performed to remove the fractured screw.
|
| |
|
Event or Problem Description
|
|
It was reported that an everest fenestrated polyaxial screw fractured post-operatively.The patient did not fuse and revision surgery was performed to remove the fractured screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d.4 was updated to reflect lot/serial number received.Device and complaint history records were reviewed and no relevant manufacturing issues or similar complaints were identified.
|
| |
|