MAUDE Adverse Event Report: EXIMO MEDICAL LTD. B-LASER ATHERECTOMY CATHETER 1.5 MM OTW; PERIPHERAL ATHERECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXIMO MEDICAL LTD. B-LASER ATHERECTOMY CATHETER 1.5 MM OTW; PERIPHERAL ATHERECTOMY CATHETER Back to Search Results
Model Number EXM-4001-0000
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 04/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported device has been returned to the manufacturer.The plant investigation is in process.Reference (b)(4).
 
Event or Problem Description
At the closure of an atherectomy procedure, it was noted an eximo 1.5 mm catheter blade had detached while inside of a patient.Attempts to remove the detached blade from within the patient were unsuccessful.The blade was not removed.The patient was reported as stable.The reported device has been returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was an auryon catheter.The catheter was forwarded to angiodynamics' manufacturer of the product for evaluation and root cause determination.Per the evaluation, the customer's reported complaint description of tip detached is confirmed.The blade tip was detached from the end of the catheter and was not returned.The tip of the catheter shows an expected pattern of use for an extended period of time.The root cause is likely related to handling damage during use of the catheter, i.E.Lasing for an extended period of time.Instructions for use (ifu) cautions against lasing at the same location for more than 10 seconds, i.E.With no catheter advancement.A review of the device history record (dhr) for catheter lot: ex017ao78 was performed to determine if any deviations related to the reported defect of the complaint.The review confirmed that the lot met all material, assembly and performance specifications.Labeling review: instructions for use is provided with the catheter device and contain the following statements: warnings: pay careful attention while using the catheter, avoid excessive force and be on alert for any potential damage.Inadvertent movement of the catheter may result in patient injury.Potential complications: distal embolization.Note: it is expected, especially with cto lesions at the cap, that the advancement rate may be slower.In any such case, and in any other occasion that the catheter does not seem to be advancing at a certain point, please follow the instructions below: a) do not to exceed 10 seconds of lasing at the same location.If you experience any difficulty to advance the auryon catheter, immediately start self-count-down.Self-count-down should start the moment you experience non-advancement of the auryon catheter.When advancement resumes, you should stop self-count-down and resume it if additional non-advancements are experienced.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
B-LASER ATHERECTOMY CATHETER 1.5 MM OTW
Common Device Name
PERIPHERAL ATHERECTOMY CATHETER
Manufacturer (Section D)
EXIMO MEDICAL LTD.
3 pekeris street
glens falls, ny 12801, rehovot 76702 03
IS  7670203
MDR Report Key10052305
Report Number1319211-2020-10005
Device Sequence Number948286
Product Code MCW
UDI-Device Identifier07290017590004
UDI-Public07290017590004
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K181642
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2022
Device Model NumberEXM-4001-0000
Device Catalogue NumberEXM40010000
Device Lot NumberEX017AO78
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2020
Supplement Date Received by Manufacturer04/23/2020
Initial Report FDA Received Date05/13/2020
Supplement Report FDA Received Date08/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-