• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION V-18 CONTROL WIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION V-18 CONTROL WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40495
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation by mfr.: unit returned with its original box and together with a non-bsc catheter, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The guidewire returned together with its original opened pouch; besides, a non-bsc device was received.The device has the body kinked approximately at 7.5 cm, 49 cm, 77.5 cm and 123 cm from the proximal end.The guidewire distal tip was bent.The polymer distal tip was damaged.No more damages were found in the device.During the inspection, it was noted that the proximal end of the polymer tip was folded on itself; the folded section measured approximately 0.4 cm.The overall length of the device is within specification.The outer diameter of the distal tip, middle of the device, and proximal section are within specification.
 
Event or Problem Description
Reportable based on device analysis completed on 07may2020.It was reported that inserting difficulties were encountered.The 99% stenosed target lesion was located in the severely tortuous and mildly calcified forearm shunt.A 150cm, 8cm poly tip v-18 control wire guidewire was selected for use in combination with a non-bsc balloon catheter.However, the guidewire had difficulty inserting into the guidewire lumen of the balloon catheter.Both devices were removed and the procedure was completed with another of same device.No patient complications nor injuries were reported.However, returned device analysis revealed guidewire coating peeled/sheared.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-18 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10052460
Report Number2134265-2020-06368
Device Sequence Number2782877
Product Code DQX
UDI-Device Identifier08714729070184
UDI-Public08714729070184
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K033742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/29/2021
Device Model Number40495
Device Catalogue Number40495
Device Lot Number0024356010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/07/2020
Initial Report FDA Received Date05/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-