• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number SDMDK1000
Device Problems Positioning Failure (1158); Migration or Expulsion of Device (1395); Application Program Problem (2880)
Patient Problem No Code Available (3191)
Event Date 04/24/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).(cancelled case) out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during the procedure, the physician had issues deploying the fiducials, each one migrated upon deploying out of the end of the marker delivery kits.When tech was loading the fiducial into the first marker delivery kit it became lodged in the proximal hub and could not advance the guide wire.The superd portion was not completed.It was not completed by other means.The patient was under general anesthesia.
 
Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Two marker delivery catheter kits in their original packaging were returned with 6 empty marker cartridges.Visual inspection noted that there was a fiducial marker stuck in the hub of one of the delivery catheters.There was an obvious deformation of the stuck marker upon removal.The supplied guide wire was ran through the delivery catheters and there was no obstruction.It was reported that the device was unable to deploy and the fiducial marker migrated.The reported issue was confirmed.The most likely root cause could not be established from the information available.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the fiducial marker was unable to deploy from the marker deliver kit and migrated inside the patient.Also, the patient procedure was aborted due to a system issue.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERDIMENSION
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
MDR Report Key10054073
MDR Text Key196571346
Report Number3004962788-2020-00027
Device Sequence Number1
Product Code JAK
UDI-Device Identifier10884521200524
UDI-Public10884521200524
Combination Product (y/n)N
PMA/PMN Number
K110093
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/30/2022
Device Model NumberSDMDK1000
Device Catalogue NumberSDMDK1000
Device Lot NumberE1634266
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2020
Date Manufacturer Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-