| Model Number |
ZCU600 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Sex/gender: unknown/ not provided.Date of event: unknown, not provided.Best estimate of date of event is between (b)(6) 2020 and (b)(6) 2020.(b)(4).An attempt has been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's left eye due to poor post-op outcome (overcorrection).There was no incision enlargement.The explanted iol was replaced with a model zcu375 20.0 diopter lens.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d10: device available for evaluation? yes.Returned to manufacturer on: 5/18/2020.Section h3: device returned to manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the returned condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|