• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 S-ROM*SLEEVE PRX ZTT, 18B-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 S-ROM*SLEEVE PRX ZTT, 18B-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 52-1483
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Tissue Damage (2104); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 11/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pcf and medical records received.After review of medical records, patient was revised to address failed left total hip replacement, aseptic loosening of left total hip replacement, and failed metal-on-metal total hip arthroplasty.Patient was with a painful left total hip, was found to have aseptic loosening of her left femoral stem, and was also found to have a metal-on-metal total hip arthroplasty.Operative notes stated that the short external rotators were identified and tagged for later repair.The femoral stem was grossly loose and was able to move it by hand.The s-rom ztt proximal sleeve and stem were well connected.The hoop extractor was used and following removal of lateral bone with a high-speed bur, the stem was removed without any bone loss.It was grossly loose.Reverse curettes were used to clean the pseudomembrane out the femoral canal.Scar and pseudocapsule was debulked.The leg lengths with disarticulation were recreated.The patient was short prior to surgery from the loose subsided stem.Doi: (b)(6) 2011.Dor: (b)(6) 2019; (left hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10056620
Report Number1818910-2020-12096
Device Sequence Number14292241
Product Code LPH
UDI-Device Identifier10603295170631
UDI-Public10603295170631
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K934412
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2016
Device Model Number52-1483
Device Catalogue Number521483
Device Lot Number3308201
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/04/2020
Supplement Date Received by Manufacturer06/28/2020
Initial Report FDA Received Date05/14/2020
Supplement Report FDA Received Date06/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
-
-