| Model Number |
PED-500-18 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a pipeline device that failed to open at the distal end.The patient was being treated for an unruptured saccular aneurysm of the left interior carotid artery.The aneurysm man diameter was 5mm and the neck diameter was 4mm.Vessel tortuosity was moderate.It was reported that all the devices for the procedure were prepared according to the manufacturer instructions for use.The pipeline was delivered but the distal segment of the stent would not open up completely.The pipeline was replaced.There was no harm or injury to the patient.Ancillary devices: phenom 27 microcatheter and phenom plus distal access catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the pipeline was initially opened in the straightened segment below the terminus then dragged back to aneurysm for deployment which had many bends.Multiple attempts were made to open the distal segment of the pipeline flex including reconstraining to the ptfe sleeve segment , vigorous loading and unloading flattened pipe segment( wagging), centering phenom 27 in middle of vessel during deployment, and repositioning the pipeline flex segment before the vessel bend.There was no unusual resistance felt during deployment of the pipeline flex during delivery of the device.
|
| |
|