• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL IRELAND LTD. INQWIRE.GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL IRELAND LTD. INQWIRE.GUIDE WIRE Back to Search Results
Model Number 00884450021576
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device has returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during incoming inspection of the guidewire packaging, an open seal was identified compromising the sterility of the wire.A new guidewire was used to complete the procedure.No patient contact or injury to report.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the transport and handling process.A search of the complaint database was performed and one similar complaint for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQWIRE.GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
MERIT MEDICAL IRELAND LTD.
parkmore business park west
galway, ballybrit
EI 
MDR Report Key10059720
Report Number9616662-2020-00009
Device Sequence Number14182260
Product Code DQX
UDI-Device Identifier00884450021576
UDI-Public00884450021576
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K133230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number00884450021576
Device Catalogue NumberIQ35F180J3/B
Device Lot NumberK1791811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2020
Initial Date Received by Manufacturer 04/20/2020
Supplement Date Received by Manufacturer06/05/2020
Initial Report FDA Received Date05/14/2020
Supplement Report FDA Received Date07/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-