• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1712K
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Swelling (2091)
Event Date 05/13/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that the customer's insulin pump had the retainer ring detached.The customer's blood glucose level was unknown at the time of the incident.There was no confirmation if the reservoir locks in place.No further details were provided.The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
Unable to perform displacement test due to detached retainer and missing reservoir tube o-ring.Test reservoir does not lock properly inside the reservoir tube due to detached retainer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10060765
MDR Text Key191101306
Report Number2032227-2020-118446
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00763000192143
UDI-Public(01)00763000192143
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712K
Device Catalogue NumberMMT-1712K
Device Lot NumberHG34TED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received06/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-