U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ B X 19.5MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ B X 19.5MM, E-PLUS Back to Search Results
Model Number 315-0B-719
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to an infection.The previous surgery and the surgery detailed in this event occurred 3.5 years apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available.The revision surgery was completed as intended and without incident.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate that the reported device was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.
 
Event or Problem Description
Revision surgery - due to possible infection, removed poly and did a thorough washout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT REVISION KNEE
Common Device Name
INSERT, REVISION EXPRT, SZ B X 19.5MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd
austin, tx
MDR Report Key10061700
Report Number1644408-2020-00392
Device Sequence Number19711441
Product Code JWH
UDI-Device Identifier00888912114028
UDI-Public(01)00888912114028
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number315-0B-719
Device Catalogue Number315-0B-719
Device Lot Number010R1005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2020
Initial Report FDA Received Date05/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
-
-