TERUMO MEDICAL CORPORATION DESTINATION PERIPHERAL GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that the destination hub was not connected to the sheath.The hub fell on the floor when the sterile package was opened.A new sheath was opened.Additional information was received on 27apr2020.The issue occurred during prepping, without the patient present.Sine the hub was not attached, it fell on the floor.There was no patient injury, medical/surgical intervention required.Additional information was received on 05may2020.The reported event was discovered upon opening the package.The procedure being performed was an angiogram of the lower extremities.The patient's condition was stable.Another sheath was opened, and the procedure was completed.Additional information was received on 06may2020.A new terumo sheath was opened, and the case was continued with successful turnouts.The patient was present when the issue occurred.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned for evaluation; however, a photograph of the actual device was returned for evaluation.Therefore, the investigation was based upon the user facility information and a photograph provided by the user facility.Visual inspection of the provided image revealed that the valve assembly was unscrewed from the sheath in the package tray.No other anomalies were noted.It is likely that the valve assembly could have unscrewed from the sheath hub at an unknown time postproduction.However, with no return of the actual device, the exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESTINATION PERIPHERAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton, md
MDR Report Key10061822
Report Number1118880-2020-00109
Device Sequence Number3514704
Product Code DYB
UDI-Device Identifier00389701008837
UDI-Public00389701008837
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172995
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model NumberN/A
Device Catalogue Number54-64501
Device Lot NumberYA22
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/23/2020
Supplement Date Received by Manufacturer06/09/2020
Initial Report FDA Received Date05/15/2020
Supplement Report FDA Received Date06/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-