• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PERI-LOC K-WIRE 1.6MM; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PERI-LOC K-WIRE 1.6MM; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 71161016
Medical Device Problem Codes Break (1069); Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an orif of left humerus, the surgeon pinned a plate to the humerus with a peri-loc k-wire to verify its placement with x-ray imaging.Upon removal of the wire, the tip of it broke and remained inside the patient.It is unknown whether surgery was delayed or not.The patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment was not returned for evaluation.Without the actual product, product evaluation could not be performed and complaint could not be confirmed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.According to clinical/medical investigation, the case reports a broken peri-loc k-wire tip that was retained in the patient.Although it was noted that x-rays were taken during the procedure, no x-ray images or clinically relevant supporting documentation was provided for inclusion in this medical investigation.Based on the limited information provided the root cause for the reported breakage could not be determined.The material of the k wire tip is 17-4 ph stainless steel and long-term implantation data is not available.The patient impact beyond possible corrosion, local irritation/discomfort, and/or migration of possibly retained non-implantable foreign body fragments cannot be determined.No further medical assessment can be rendered at this time.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will not be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERI-LOC K-WIRE 1.6MM
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10064201
Report Number1020279-2020-01765
Device Sequence Number10906123
Product Code HTY
UDI-Device Identifier03596010179708
UDI-Public3596010179708
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71161016
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/18/2020
Supplement Date Received by Manufacturer01/19/2021
Initial Report FDA Received Date05/17/2020
Supplement Report FDA Received Date01/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age23 YR
Patient Weight53
-
-