| Model Number |
MMT-751LNAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the button's were hard to press and insulin pump rejected new batteries.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with all buttons responding properly.No damage on keypad assembly.No unlocked lcd keypad connector.Device received with all operating currents within specification and passed a21 error test and displacement test.No unexpected failed battery alarm anomalies noted.(b)(4).
|
| |
|