| Model Number |
03.037.120 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure on (b)(6) 2020, the driving cap would not seat fully on the threaded hammer guide connector.The surgeon used the assembly in the partial seated position to insert the nail.The procedure was successfully completed.There was no patient consequence and no surgical delay reported.Concomitant devices reported: tfna nail (part number unknown, lot unknown, quantity 1), threaded hammer guide connector (part number 03.037.120, lot l735215, quantity 1), driving cap/threaded (part number 03.010.523, lot l759641, quantity 1).This report involves 1 threaded hammer guide connector.This is report 3 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part: 03.037.120-us, lot: 9113148, manufacturing site: hägendorf, release to warehouse date: 12.Sept.2014.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the threaded hammer guide connector (part # 03.037.120 / lot # 9113148) was received at us customer quality (cq).The device showed no signs of visual damage.Functional test: functional testing was performed with the returned driving cap/threaded.Due to the observed damage on the driving cap the devices weren¿t able to properly assembly.Due to the stripped condition of the driving caps threads, the devices do not mate properly.The hammer guide connector did not contribute to the complaint condition, thus the complaint was not confirmed for the hammer guide.Can the complaint be replicated with the returned device(s)? yes; but the issue was solely caused by the damaged driving cap.The hammer guide did not contribute to the complaint condition.Dimensional inspection: dimensional inspection was not performed since no defect was found with the hammer guide.Document/specification review: drawing(s) reviewed: (current & manufactured revisions).Conclusion: the overall complaint was not confirmed for the received threaded hammer guide connector as the device assembly issue was caused by the stripped condition of the mating threaded cap.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|