WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM / TI CANN FRN / PF 380MM / LEFT - STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.033.139S |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/ investigation but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/ or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020, the patient had bilateral midshaft femur fractures.On (b)(6) 2020 a surgery was performed with bilateral, antegrade intramedullary nails.The surgeon started on the left femur first, the proximal femur was opened and the canal was reamed to 12.5mm and proximally to 14mm to accommodate an 11 mm x 380mm left femoral recon nail.The nail was inserted using standard technique and approximately two (2) to three (3) inches from final placement depth, the nail would no longer advance.The surgeon used typical force to advance the nail to its final position, but it would not go any further.A series of ap and lateral x-rays were taken to make sure nothing was impeding the nail path, and nothing was found.The surgeons decided to remove the nail which took an extraordinary amount of force to remove and added time to the procedure.The nail was removed, and the surgeon decided to ream the canal to a larger diameter.The canal was then reamed to 13.5 mm and the nail was re-inserted.The nail stopped advancing in the same spot, as it had the first time.Even with heavy impaction, it would not advance any further.The surgeons decided to remove the nail which again took an extraordinary amount of force and time to remove.The surgeon decided to switch to a different implant and chose an 11 mm x 380 mm retrograde/ antegrade femoral nail.The new nail went in without incident.The right femur was also treated with an 11 mm x 380 mm retrograde/ antegrade femoral nail.It also went in without incident.There was approximately a thirty (30) minute surgical delay.The procedure was completed successfully.The patient was reported as stable, after the procedure.Concomitant devices reported: reamer (part number unknown, lot unknown, quantity 1).This report involves one (1) 11mm / ti cann frn / pf 380mm / left ¿ sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: b4: alert date change from 05/05/2020 to 05/04/2020.H3, h6: investigation summary : product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.Device history lot: part: 04.033.139s, lot: 4l45206, manufacturing site: bettlach, release to warehouse date: sep.01, 2016, expiry date: april 30, 2029.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|