| Model Number |
M003SFD040300 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); High Blood Pressure/ Hypertension (1908); Rupture (2208)
|
| Date of Event |
05/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The subject device remained inside patient.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure the subject device (stent) was deployed prematurely in the wrong location.The subject device stent was fully deployed from middle cerebral artery (mca) to cavernous internal carotid artery (ica) and controlled bleeding to an extent but landed further distal than anticipated.The physician tried to drag the subject stent proximal into the ica and the patient¿s blood pressure spiked up which caused an aneurysm rupture near distal segment of supraclinoid ica.The subject device was unable to be removed once fully deployed so left in place and decided not to coil.The patient was administered with the medication in response to the aneurysm rupture.Patient had an evd (external ventricular drain) placed immediately after the procedure.Patient¿s current status is sick but stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.As per the available information, the device deployed prematurely in the wrong location, too far distal.The physician tried to drag the stent back and the patient's blood pressure spiked and caused an aneurysm rupture.As the product was not returned and review and analysis of all available information fails to indicate an assignable cause or probable assignable cause for the stent premature deployment, and it cannot be definitively determined if the stent premature deployment caused or contributed to the aneurysm rupture and patient complications, an assignable cause of undeterminable will be assigned to this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure the subject device (stent) was deployed prematurely in the wrong location.The subject device stent was fully deployed from middle cerebral artery (mca) to cavernous internal carotid artery (ica) and controlled bleeding to an extent but landed further distal than anticipated.The physician tried to drag the subject stent proximal into the ica and the patient¿s blood pressure spiked up which caused an aneurysm rupture near distal segment of supraclinoid ica.The subject device was unable to be removed once fully deployed so left in place and decided not to coil.The patient was administered with the medication in response to the aneurysm rupture.Patient had an evd (external ventricular drain) placed immediately after the procedure.Patient¿s current status is sick but stable.
|
| |
|