The reported failure "tube disconnected from the outlet connector "occurred during use.The affected oxygenator is not available for further investigation by the manufacturer.Customer did not save the device.A review for similar complaints to be investigated already was performed and no similar complaints were found.According to the ssu dated on (b)(6)2020 the tubes were not secured with ties on the outlet connector.Thus the failure could not be confirmed.According to the life cycle engineering (lce) consumables the tolerance range on the 701067859 beq-hmod70000-usa #squadrox-id ad.O.Fil.(screw connector) has not been changed compared to the oxygenator with glue connector.All external dimensions of the connector (relevant with regard to the connection) have not been changed.(see email in the attachment) as stated in the complaint the used tubing set was from the manufacturer livanova.According to the ssu liva nova is also informed about the incident.The risk assessment quadrox-i small adult/adult, quadrox-id adult dms # 1464420 v15 chapter r6.1.1.1,r6.1.1.2, r6.1.1.3, r6.1.1.4, r6.1.1.5 and r6.1.1.6 was reviewed on 2020-06-10 with the following outcome: the most possible causes for the reported failure "tube disconnected from the inlet connector" could be determined as: * lack of attention on device handling * inappropriate fixation the reported failure "tube disconnected from the inlet connector" occurred during use and could not be confirmed.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending h3 other text : 4115.
|