ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ACU0T0
Device Problem Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Date 05/07/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A health professional reported that during an intraocular lens (iol) implant surgery, the lens was stuck in the injector.There was patient contact with the device.A backup device was used to complete the case.There was no patient harm.
 
Manufacturer Narrative
The device with the lens was returned.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger has been advanced over the lens to mid-nozzle.The lens is still in the loading area.The trailing haptic is folded in on the anterior optic.Product history records were reviewed and documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if the qualified product was used.The root cause could not be determined.A plunger override was observed.The plunger position advanced to mid-nozzle with the lens still in the loading area may indicate the plunger was advanced too rapidly.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key10067150
MDR Text Key191606748
Report Number1119421-2020-00779
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model NumberACU0T0
Device Lot Number12659661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Date Manufacturer Received07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age72 YR
-
-