| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
05/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device received with all buttons responding properly.No damage or moisture damage on keypad assembly.No unlocked lcd keypad connector.Device passed self test.
|
| |
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that buttons of the insulin pump were not responding.It would not work with any buttons and she pressed on tried down arrow button and it was nothing worked.Customer also received failed battery alarm.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|