Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, pre operatively, on a laparoscopic total gastrectomy, at the first firing, when the charger was being inserted by the handle was checked, the red indicator was noted to be illuminated.When the connection part was being contacted by handle and the charger was checked, the event like pin melting was noted.The procedure was completed with another device.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: "medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the clinical power handle returned was identified that the battery cells had vented.A review of the system logs showed there were multiple thermistor error messages at the end of the event log.It was reported that the device showed signs of melting.The reported issue could not be confirmed.The most likely root cause could not be established from the information available.It was also reported that the battery charger light was red indicating a device issue.The reported issue was confirmed.The most likely root cause was not traced to the device.This battery electrolyte most likely contaminated the charging bay; this contamination prevented the charger pogo pins from functioning as intended.The electrolyte is conductive, and would therefore create false readings on the thermistor line, leading to the red error lights identified in the software event log.This would prevent the powered handle from charging.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|