• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT2H04
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the posterior communicating artery (pcom) using penumbra smart coils, non-penumbra microcatheter and a guidewire.It was noted that the patient anatomy was tortuous.During the procedure, the physician successfully implanted two smart coils in the target vessel using the microcatheter and guidewire.Then, the physician attempted to advance the next smart coil within its introducer sheath; however, the smart coil became stuck within its introducer sheath.Therefore, the smart coil was removed.The procedure was completed using three additional smart coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the posterior communicating artery (pcom) using penumbra smart coils, non-penumbra microcatheter and a guidewire.It was noted that the patient anatomy was tortuous.During the procedure, the physician successfully implanted two smart coils in the target vessel using the microcatheter and guidewire.Then, the physician attempted to advance the next smart coil within its introducer sheath; however, the smart coil became stuck after advancing a few millimeters within its introducer sheath.Therefore, the smart coil was removed.The procedure was completed using three additional smart coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section is being updated based on additional information provided by a penumbra distributor on 05/22/2020: section b.Box 5.Describe event or problem results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 11.0, 74.0, and 127.0 cm from the proximal end.The pusher assembly mid-joint was proximal to the introducer sheath friction lock.The embolization coil was intact with the pusher assembly.Conclusions: evaluation of the returned smart coil revealed that the pusher assembly mid-joint was proximal to the introducer sheath friction lock.If this occurs, resistance may be encountered upon advancement, and the smart coil may not advance within its introducer sheath.If the device is forcefully advanced against this resistance, damage such as kinks in the pusher assembly may result.During the functional test, the smart coil was able to be advanced from its returned position and out of its introducer sheath with resistance.The smart coil was then advanced through a demonstration microcatheter with resistance due to the kinks in the pusher assembly.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10071183
Report Number3005168196-2020-00661
Device Sequence Number3753327
Product Code HCG
UDI-Device Identifier00814548015323
UDI-Public00814548015323
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2023
Device Model Number400SMTXSFT2H04
Device Catalogue Number400SMTXSFT2H04
Device Lot NumberF86444
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/29/2020
Supplement Date Received by Manufacturer05/22/2020
Initial Report FDA Received Date05/19/2020
Supplement Report FDA Received Date06/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-