• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. X-ALT HIGHLY HIP; LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 36MM ID, MP7

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. X-ALT HIGHLY HIP; LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 36MM ID, MP7 Back to Search Results
Model Number 932-36-252
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to an infection.The previous surgery and the surgery detailed in this event occurred 11.6 years apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available.The revision surgery was completed as intended and without incident.The device was kept at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate that the reported device was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to an infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-ALT HIGHLY HIP
Common Device Name
LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 36MM ID, MP7
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd
austin, tx
MDR Report Key10071257
Report Number1644408-2020-00394
Device Sequence Number11706237
Product Code OQG
UDI-Device Identifier00888912100076
UDI-Public(01)00888912100076
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130365
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2013
Device Model Number932-36-252
Device Catalogue Number932-36-252
Device Lot Number53910301
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/24/2020
Supplement Date Received by Manufacturer05/14/2020
Initial Report FDA Received Date05/19/2020
Supplement Report FDA Received Date05/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-035, LOT 53919817; 497-36-000, LOT 53890352
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-