| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned device identified that the dovetail feature was flared out and compressed.The device also exhibits damage in various areas on distal surface probably that happened during insertion.Device history record was reviewed and no discrepancies were found.The root cause is attributed to use error.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty, the tibial insert was warped during insertion and could not be implanted.No adverse events have been reported as a result of the malfunction.
|
| |
|