• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HUBER PLUS 20G X 0.75" NEEDLELESS Y SITE SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HUBER PLUS 20G X 0.75" NEEDLELESS Y SITE SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Patient Involvement (2645)
Event Date 04/28/2020
Event Type  malfunction  
Manufacturer Narrative
The information provided by bd represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd. The device has not been returned to the manufacturer for evaluation. A lot history review (lhr) of reds1094 showed 68 other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the plastic packaging is deformed, which leads to the loss of sterility of certain needles. It was state that the valve on the y-site is also deformed causing leaking. It was stated this occurred with 40 needles. This report addresses 35 of 40.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHUBER PLUS 20G X 0.75" NEEDLELESS Y SITE
Type of DeviceSET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key10072912
MDR Text Key191891254
Report Number3006260740-2020-01801
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K993848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 05/19/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/30/2022
Device Model NumberN/A
Device Catalogue Number012034NY
Device Lot NumberREDS1094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location Hospital
Date Manufacturer Received04/28/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-