MAUDE Adverse Event Report: BARD ACCESS SYSTEMS TRI-FUNNEL REPL GAST 24F; FEEDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS TRI-FUNNEL REPL GAST 24F; FEEDING DEVICE Back to Search Results
Model Number 000724
Medical Device Problem Codes Contamination /Decontamination Problem (2895); Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the device is pending.The investigation of the reported event is currently underway.(expiry date: 06/2023).
 
Event or Problem Description
It was reported that fibers were allegedly found on the end of the tube.There was no reported patient involvement.
 
Event or Problem Description
It was reported that fibers were allegedly found on the end of the tube.There was no reported patient involvement.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: one tri-funnel replacement gastrostomy tube 24f was returned for evaluation.The gross visual observed are the sample appeared to be clean.Unknown fibers were noted to the distal end of the gastrostomy tube.The microscopic visual observed the unknown fibers were noted on the outside of the balloon.Functional testing, tactile evaluation and other evaluation/testing was not performed due to the nature of this complaint.The investigation is confirmed for contamination problem, as unknown fibers were found outside of the balloon.A definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date: 06/2023),g4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-FUNNEL REPL GAST 24F
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10073172
Report Number3006260740-2020-01810
Device Sequence Number18976189
Product Code KNT
UDI-Device Identifier00801741037177
UDI-Public(01)00801741037177
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000724
Device Catalogue Number000724
Device Lot NumberNGDW1641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Initial Date Received by Manufacturer 04/27/2020
Supplement Date Received by Manufacturer10/21/2020
Initial Report FDA Received Date05/19/2020
Supplement Report FDA Received Date10/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-