| Model Number |
OPO80 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown as it was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
A surgery center reported observing holes worn into the tyvek lid of a compact intutiv pack disposable tubing set model opo80.The surgery center had concerns the packaging had been comprised and the tubing pack was no longer sterile.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10.Device available for evaluation? yes.Returned to manufacturer on: 5/29/2020.Section h3.Device returned to manufacturer? yes.Device evaluation: the tubing pack was returned to evaluation.A visual evaluation was performed.The visual assessment observed a pinhole on the tyvek lid.The pin hole like tear was observed on the tyvek lid.Manufacturing record review: per manufacturing records review report, no related non-conformity or deviations were issued during manufacturing the tubing pack lot# 60169162.All devices met material, assembly and performance specifications at the time of product released.Investigations: the failure mode of ¿pinhole in the packaging¿ was not replicated through mechanical stress inflicted by simulated standard shipping and handling conditions, however, was able to be replicated when a single opo80 tubing pack was dropped from a height of 48 inches.A potential contributor for the ¿pinhole in the package¿ may be related to the opo80 tubing pack being able to move within the tyvek packaging and during its inappropriately handling by the user such as unintentional dropping of the tubing pack tray on the floor.Device labeling does contain instructions to not use the device if the package is damaged.As a proactive and precautionary measure, the current packaging will be reconfigured (and validated) to increase its durability.Johnson and johnson surgical will continue monitor this type of complaints.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.H3 other text : placeholder.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|