MAUDE Adverse Event Report: RICHARD WOLF GMBH NEEDLE PUSHER WL 210MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RICHARD WOLF GMBH NEEDLE PUSHER WL 210MM Back to Search Results
Model Number 826750
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation of the device is in progress.The device is currently being returned by the user to rw (b)(4), and a follow-up report will be submitted as soon as the investigation is completed.
 
Event or Problem Description
Richard wolf (b)(4) received the following information on may 07, 2020: the operation " microlaryngoscopy and laterofixation of the vocal cord" could not be performed due to the malfunction.Change to surgery with co2 laser.The needle is very difficult to push out of the needle guide, therefore the surgeon has to apply high pressure.Furthermore, the locking of the holder does not work when the needle is retracted.A correct laterofixation could not be performed.Change to surgery with co2 laser.There was no danger to the patient or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE PUSHER WL 210MM
Common Device Name
NEEDLE PUSHER WL 210MM
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer street 32
knittlingen, 75438
GM  75438
MDR Report Key10073483
Report Number9611102-2020-00006
Device Sequence Number2911356
Product Code LRC
UDI-Device Identifier04055207039199
UDI-Public04055207039199
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number826750
Device Catalogue Number826750
Device Lot Number1019440
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-