| Model Number |
21-2766-24 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
04/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received indicating that needle stick occurred with use of a smiths medical deltec gripper plus safety needle.It was reported that the safety arm would not lock, and the needle sprung out of it.There were no reported adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned device was received for evaluation.During the evaluation of the device we observed that safety mechanism was already activated, and the needle was out of the base.The customer reported condition was confirmed.Problem source is unknown.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|