• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00545160
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was used in papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, there was a break in the cutting wire when it was pulled.Nothing detached and fell into the patient.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: device code 1069 captures the reportable event of wire break.Block h10: the autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire of the device was found detached, bent and blackened.It was also observed that the break in the wire was not smooth.A functional evaluation could no be performed due to the condition of the device.No other issues with the device were noted.The reported event was confirmed.According to the product analysis, the cutting wire of the device was found detached, bent and blackened.Based on the condition of the device, the observed damage may have been caused by excessive voltage or energizing the device when not in contact with tissue.Based on review and analysis of all available information, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was used in papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, there was a break in the cutting wire when it was pulled.Nothing detached and fell into the patient.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOTOME RX 49
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10075720
Report Number3005099803-2020-01894
Device Sequence Number2848207
Product Code KNS
UDI-Device Identifier08714729415152
UDI-Public08714729415152
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2022
Device Model NumberM00545160
Device Catalogue Number4516
Device Lot Number0024613551
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2020
Initial Date Received by Manufacturer 04/22/2020
Supplement Date Received by Manufacturer05/21/2020
Initial Report FDA Received Date05/20/2020
Supplement Report FDA Received Date06/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight64
-
-