| Model Number |
R SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's report was observed during initial testing; however after disassembling the device to determine root cause, the customer's report was no longer seen.The device was put through extensive testing without duplicating the report.The pace/defib engine board was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" and "defib fault 76" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|