| Model Number |
28160GE |
| Medical Device Problem Code |
Solder Joint Fracture (2324)
|
| Health Effect - Clinical Code |
Brain Injury (2219)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was recently returned to the manufacturer.Per their initial inspection: a broken weld joint on the endoscope locking mechanism was seen, and there are several spots of corrosion - very strong inside the endoscope locking mechanism.Part is from november 2006.
|
| |
|
Event or Problem Description
|
|
As per a report we received from the factory in (b)(4) regarding an event that took place in (b)(6): handling difficult during operation; weld problem on the three-way stop cock.Damage to the left side of the brain postoperative.
|
| |
|