| Model Number |
0747031650 |
| Medical Device Problem Codes |
Contamination /Decontamination Problem (2895); Contamination of Device Ingredient or Reagent (2901)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that debris remained inside the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that debris remained inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
*note: the investigation was previously closed based on product not received; however, the product has now been physically received at stryker endoscopy, usa and the investigation has been re-opened.Investigation as follows is now based on product received.Alleged failure: noticed ¿fuzzy¿ quality to screen image, not specific to any one case.Then they identified the actual bioburden issue when it was sent out to 3rd-party repair salesforce case number (b)(4).Additional information received states that ¿flecks of cooked bioburden came out and into the patient¿.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be third party repair and/or improper sterilization methods.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|