| Model Number |
0747031650 |
| Medical Device Problem Code |
Contamination of Device Ingredient or Reagent (2901)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that debris remained inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: noticed ¿fuzzy¿ quality to screen image, not specific to any one case.Then they identified the actual bioburden issue when it was sent out to 3rd-party repair the probable root cause/s could be third party repair and/or improper sterilization methods.
|
| |
|
Event or Problem Description
|
|
It was reported that debris remained inside the patient.
|
| |
|