Catalog Number 71453121
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/24/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented an infection on the knee.A revision surgery and a wash out were performed to correct the issue.Poly was swapped.
 
Additional Manufacturer Narrative
It was reported that the patient presented an infection on the knee.A revision surgery and a wash out were performed to correct the issue.Poly was swapped.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.As device information was not made available, device history record review cannot be completed.There is no information that would suggest the device failed to meet specifications.A relationship, if any, between the device and the reported incident could not be corroborated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Based on this investigation, the need for corrective action is not indicated.Infection, a potential complication associated with any surgery, can occur and possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN CR HIGH FLEX XLPE SZ 5-6 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10077176
Report Number1020279-2020-01812
Device Sequence Number11602416
Product Code JWH
UDI-Device Identifier00885556033982
UDI-Public885556033982
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71453121
Initial Date Received by Manufacturer 04/24/2020
Supplement Date Received by Manufacturer01/25/2021
Initial Report FDA Received Date05/20/2020
Supplement Report FDA Received Date02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-