|
SMITH & NEPHEW, INC. LEGION PS OXIN FEM SZ6 LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META
|
Back to Search Results |
|
| Catalog Number |
71421216 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
12/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the clinical subject experienced decreased range of motion on the study knee.The event was treated via surgery and is considered resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: it was reported that the clinical subject experienced decreased range of motion on the study knee.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.Based on this investigation, the need for corrective action is not indicated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.A medical analysis noted that the patient developed a painful secondary patellofemoral osteoarthritis.Patella tangential x-ray views in the time course are provided and show narrow joint space and change in angle of the patella as well as bone growth on the borders of the patella.Patient was advised to undergo secondary patellar resurfacing but has not made an appointment for that surgery.Instead, approximately 3 months post implantation the patient underwent an ¿arthroscopic arthrolysis¿ and the event is now documented as resolved.The impact to the patient beyond the second surgery cannot be determined.The root cause of the decrease range of motion 4 months post op could be scaring after the procedure.It does not appear to be a failure of the devices implanted.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: it was reported that the clinical subject experienced decreased range of motion on the study knee.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.Based on this investigation, the need for corrective action is not indicated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.A medical analysis noted that the patient developed a painful secondary patellofemoral osteoarthritis.Patella tangential x-ray views in the time course are provided and show narrow joint space and change in angle of the patella as well as bone growth on the borders of the patella.Patient was advised to undergo secondary patellar resurfacing but has not made an appointment for that surgery.Instead, approximately 3 months post implantation the patient underwent an ¿arthroscopic arthrolysis¿ and the event is now documented as resolved.The impact to the patient beyond the second surgery cannot be determined.The root cause of the decrease range of motion 4 months post op could be scaring after the procedure.It does not appear to be a failure of the devices implanted.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|