• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551LNAB
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that insulin pump had motor error alarm.Customer reported that the drive support cap was normal.Customer was able to clear but not able to rewind the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with a35 alarmed during rewind due to motor encoder signal out of phase.Unable to perform the rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test or verify motor error alarm due to a35 alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551LNAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10078111
Report Number2032227-2020-120691
Device Sequence Number17110423
Product Code OZO
UDI-Device Identifier00643169507692
UDI-Public(01)00643169507692
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551LNAB
Device Catalogue NumberMMT-551LNAB
Device Lot NumberA4551LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Initial Date Received by Manufacturer 05/16/2020
Supplement Date Received by Manufacturer10/12/2020
Initial Report FDA Received Date05/21/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight189
-
-