| Catalog Number |
H93863 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that four (4) trifurcated fluid transfer tube sets had heat seals on the packaging that came apart.This issue was identified during setup and prior to patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: the lot was manufactured from september 07, 2018, september 11, 2018.H10: six (6) actual samples were received for evaluation.Unaided visual inspection was performed which observed that the bottom primary packaging were opened of all samples.The reported condition was verified.The cause of the condition was not determined.A functional testing was not performed for this complaint.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|