• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 9MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 9MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71453231
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Date of Event 12/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the clinical subject experienced decreased range of motion on the study knee.The event was treated via surgery and is considered resolved.
 
Additional Manufacturer Narrative
It was reported that the clinical subject experienced decreased range of motion on the study knee.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.Based on this investigation, the need for corrective action is not indicated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.A medical analysis noted that the patient developed a painful secondary patellofemoral osteoarthritis.Patella tangential x-ray views in the time course are provided and show narrow joint space and change in angle of the patella as well as bone growth on the borders of the patella.Patient was advised to undergo secondary patellar resurfacing but has not made an appointment for that surgery.Instead, approximately 3 months post implantation the patient underwent an ¿arthroscopic arthrolysis¿ and the event is now documented as resolved.The impact to the patient beyond the second surgery cannot be determined.The root cause of the decrease range of motion 4 months post op could be scaring after the procedure.It does not appear to be a failure of the devices implanted.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 7-8 9MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10080671
Report Number1020279-2020-01829
Device Sequence Number17091794
Product Code JWH
UDI-Device Identifier03596010206299
UDI-Public3596010206299
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/22/2025
Device Model Number71453231
Device Catalogue Number71420172
Device Lot Number15KT60692
Initial Date Received by Manufacturer 04/27/2020
Supplement Date Received by Manufacturer04/27/2020
12/15/2020
Initial Report FDA Received Date05/21/2020
Supplement Report FDA Received Date06/09/2020
01/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight110
-
-