| Model Number |
H403023 |
| Medical Device Problem Code |
Environmental Compatibility Problem (2929)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference: 2184149-2020-00056, 2184149-2020-00057.During a procedure, a saline bag ruptured causing water damage to devices.The procedure was cancelled.The claris ep4 stimulator had evidence of saline fluid damage and had fluid inside the unit.The signals were intermittent and could not be repaired.The db9 serial to fiber media converter and the catheter input module 1¿cm1 pin box both had water damage.There was no harm to the patient or the staff.
|
| |
|
Additional Manufacturer Narrative
|
|
The root cause of the investigation is inconclusive since the device was not returned for analysis.Based on communication with the user the root cause of the event may be traced to liquid intrusion from a ruptured saline bag.
|
| |
|