| Model Number |
G34063 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is still pending.A follow-up report will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The rubber bands were so loosen so they could not band the hemorrhoids tightly.Physician used another new hbml-4-tri, but he felt that the banding was not as tight as the past.
|
| |
|
Additional Manufacturer Narrative
|
|
The hmbl-4-tri device of lot number cf1643730 involved in this complaint was not returned for evaluation.With the information provided, a document based investigation was conducted.Prior to distribution hmbl-4-tri devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for hmbl-4-tri of lot number cf1643730 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number cf1643730.As per the instructions for use, users are advised of the following: "if the package is opened or damaged when received, do not use.Visually inspect with particular attention to joints, cracks or breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ the ifu states that ¿ maintain suction of haemorrhoid by keeping suction port covered.Deploy band by slowly rotating suction spool downwards until tension is released.¿ ¿uncover suction port of ligator to relieve suction on haemorrhoid.This will allow ligated haemorrhoid to be released from tip of device.¿ there is no evidence to suggest that the customer did not follow the instructions for use.A definitive root cause could not be determined from the available information.A possible root cause could be attributed to user technique; if suction was not released or too much tissue was sucked into the tip of the device, this may have caused the bands to slip off the haemorrhoid as the tissue may not have bulged around the band holding the bands in place, or if the user did not begin at the most proximal location from the anal sphincter and proceed distally resulting in passing the shortshot over a previously placed band this may dislodge the band.Customer complaint is confirmed based on customer testimony.According to the information reported, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for emerging trends.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being submitted due to the completion of the investigation.The rubber bands were so loosen so they could not band the hemorrhoids tightly.Physician used another new hbml-4-tri, but he felt that the banding was not as tight as the past.
|
| |
|