• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-475-35
Medical Device Problem Codes Break (1069); Positioning Failure (1158)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The pipeline flex braid implanted in the patient.The broken segment of the pushwire remains in the patient and rest of pushwire was discarded.The device will not be returned for evaluation; therefore, product analysis cannot be performed.The device was not returned; therefore, the reported event could not be confirmed.The cause of the event cannot be conclusively determined from the provided information.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that pushwire of pipeline flex broke during the procedure.The patient underwent embolization treatment for a medium unruptured saccular internal carotid artery (ica) aneurysm located in cavernous segment.Measuring 18mmx4mm, landing zone distal 4.9mm proximal 4.8mm.The vessel was observed moderately tortuous.It was reported that due to a narrowing of the cervical carotid, the physician gave nitro.There may have been a small dissection.It was thought of not doing the case.He proceeded.Initially went in with a 4.75x25 pipeline and it was nearly all the way deployed, but we had concern that when trying to run the medtronic catheter through it might pop in and drop it into the aneurysm.Thus, we pulled that out and chose a 4.75 x 35.That was much wagging and receiving and centering, but all in all though pipeline was deploying well as.We were approaching the point of no return marker bands when the catheter curled up like a candy cane.Our fix was to maneuver the medtronic distal access support catheter over the catheter thus straightening out the catheter and deploy the rest of the pipeline in the distal access support catheter.This actually worked well, and the pipeline was now fully deployed with good wall apposition.It was at this point when we try to advance the catheter through the pipeline that we noticed a pipeline wire had separated.The physician took a medtronic revascularization device 6x20 proximal to the deployed pipeline to try to nab the wire from there.Unsuccessful.The physician then tried a 4 x 20 revascularization device to nab from the distal part of the pipeline in the m1.Unsuccessful.Tried a snare retrieval device but unsuccessful.Eventually line for enterprise stance the entire length of the remaining wire to tackle against the vessel wall.Good results.Patient doing fine.There were not any patient symptoms or complications associated with this event.The broken segment left in the patient¿s cavernous ica to m1.The angiographic result post procedure was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10083482
Report Number2029214-2020-00480
Device Sequence Number16839953
Product Code OUT
UDI-Device Identifier00847536016835
UDI-Public00847536016835
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2022
Device Model NumberPED-475-35
Device Lot NumberA861603
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2020
Initial Report FDA Received Date05/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
-
-