| Model Number |
3P36-35 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer stated that an initial architect afp result of 97.70 ng/ml was generated for a patient sample (id (b)(6)) that retested at 7.97/8.59 ng/ml (id 10184).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No customer returns were available.Historical quality metrics were reviewed and no adverse trend was identified for the customer's issue.Complaint searches determined that there is normal complaint activity for the likely cause lot.Labeling was reviewed and found to be adequate.Historical performance of the reagent lot was evaluated using world wide data from abbott link.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for this lot is within the established control limits which indicates that the lot is performing acceptably and comparable to other lots in the field.Based on the available information, no product deficiency was identified.
|
| |
|