The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms a review of the case report forms did not provide any insight into the root cause of the reported knee stiffness.No x-ray images or information on post-operative instructions were provided.Per the crf the reported left knee stiffness has been documented as resolved.Since no further harm is anticipated, no further clinical/medical assessment is warranted at this time.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes could be alignment, fit/size of device used or wear.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
|