Model Number CM-925
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) received three devices and three sealed with the appropriate packaging.The visual inspection of the opened sample noted three sutures and three needles.Two of the needles were returned disengaged from the suture.With microscopic evaluation, normal crimp and swage marks were noted on each needle.A tactile and microscopic inspection of the sealed sutures was performed with no abnormalities.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a caesarian section, when the device was pulled from the package, the needle popped off even when the device was on the swedge.Another device had the same issue.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: a3, g4.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key10084323
Report Number9612501-2020-00751
Device Sequence Number11471970
Product Code GAM
UDI-Device Identifier10884521032187
UDI-Public10884521032187
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model NumberCM-925
Device Catalogue NumberCM-925
Device Lot NumberD9G2293Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2020
Initial Date Received by Manufacturer 05/07/2020
Supplement Date Received by Manufacturer05/22/2020
Initial Report FDA Received Date05/22/2020
Supplement Report FDA Received Date06/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No