BIOMET UK LTD. CER BIOLOXD MOD HD 36MM +3 NK; HIP ARTHROPLASTY Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Wound Dehiscence (1154); Joint Dislocation (2374)
Date of Event 03/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it has been discarded.Medical product: liner standard 3.5 mm offset 36 mm i.D.For use with 56 mm o.D.Shell, catalog #: 00630505636, lot #: 64588657.Medical product: bone screw self-tapping 6.5 mm dia.35 mm length, catalog #: 00625006535, lot #: 64323444.Medical product: bone screw self-tapping 6.5 mm dia.25 mm length, catalog #: 00625006525, lot #: 63010467.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent a right tha.Subsequently, a revision surgery was performed due to dislocation and wound dehiscence.
 
Additional Manufacturer Narrative
(b)(4).Summary: as the product has not been received, the investigation was limited to the information provided.In addition, we have not been provided with any supporting documentation which could provide additional information.A review of the manufacturing history records confirms no abnormalities or deviations reported.The sterilisation certificate was reviewed and confirmed that product is sterilised within the specification range.A review of the complaint database over the last 3 years has found no similar complaints reported with the item 12-115122.Without the opportunity to examine the complaint product, root cause cannot be determined due to insufficient information.Risk assessment: linked complaint managed by zb warsaw (b)(4) has reviewed surgical notes provided.It is noted that a contributing factor of the issue that led to a revision is that the patient lost balance and fell while trying to dry feet getting out of the shower.The reason for the patients fall and lack of balance was not established in the linked complaint.Since the root cause of the reported event could not be determined with the information provided to date a specific hazard within the i/o table could not be selected for assessment.However, the harm of dislocation as a potential outcome is documented for a number of hazards and is considered as a severity score of 3 - moderate, which is defined as per the severity table within the rmr as prescribed medical or surgical intervention to preclude permanent impairment.Therefore the outcome of this complaint is considered to be in line with the anticipated outcome of dislocation.No corrective or preventive actions are deemed necessary at this time.If any additional information becomes available, then the complaint will be reopened and investigated thoroughly.
 
Event or Problem Description
It was reported that a patient underwent a right tha on march 9, 2020 by (b)(6).Subsequently, the patient was revised due to dislocation and wound dehiscence on march 29, 2020 by (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CER BIOLOXD MOD HD 36MM +3 NK
Common Device Name
HIP ARTHROPLASTY
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key10086150
Report Number3002806535-2020-00252
Device Sequence Number16760533
Product Code LZO
UDI-Device Identifier00887868248498
UDI-Public00887868248498
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K061312
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/03/2020
Device Model NumberN/A
Device Catalogue Number12-115122
Device Lot Number2986019
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/29/2020
Supplement Date Received by Manufacturer08/12/2020
Initial Report FDA Received Date05/26/2020
Supplement Report FDA Received Date08/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight89
-
-