| Model Number |
G34063 |
| Medical Device Problem Codes |
Patient-Device Incompatibility (2682); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
User sucked the hemorrhoids in place and rotated the spool releasing the band, but the band has no elasticity.Band didn't shrink.Above is the description from (b)(4), during the same procedure user then took another lot# of same product and checked the function before use while found out the band didn't shrink.No use on patient.User changed another brand of similar device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.".
|
| |
|
Event or Problem Description
|
|
This is a cancellation report.The investigation was completed on 03-nov-2020 and this complaint has been categorized as negative feedback and therefore, does not meet the reporting criteria as per report as per fda guidelines ¿medical device reporting for manufacturers (2016).Please refer to g and h for details of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
11 x hmbl-4-tri unused devices were returned and evaluated on (b)(6)2020.The device used during procedure was not returned a definitive root cause could not be determined from the available information.This complaint has been deemed to be negative feedback as per engineering input the simulated test that the user employed on the bands was not reflective of operation conditions."the bands have optimum elastic properties at body temperature.They are designed to recover to their original size at body temperature.Benchtop deployment is not representative of this".Therefore the video attached of the bands being deployed onto an open hand is not a good simulation of their use in the body.The band was not brought up to body temperature in this test manner.File no longer meets the fda reporting requirements as a malfunction report as the malfunction reported was not representative of actual use conditions and has been deemed negative feedback.
|
| |
|