| Model Number |
87092 |
| Medical Device Problem Codes |
Contamination (1120); Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that a woven 7f diagnostic electrode catheter was selected for a ventricular tachycardia (vt) ablation procedure.It was noted that visually the catheter coating was peeling and there where contaminates on the catheter.The catheter was not used and the procedure was competed with another woven catheter with no patent complications being reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a woven 7f diagnostic electrode catheter was selected for a ventricular tachycardia ablation procedure.However it was noted that the catheter had a peeling feeling and contaminates on the catheter.The catheter was not used and the procedure was competed with another woven catheter with no patent complications being reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned with the aligation of a peeling feeling and contaminates on the catheter.The visual inspection of the device revealed that the electrodes were smooth and no sharp edges.The returned device showed no evidence of the reported allegation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|